Technical Characteristics:
◆Compliant with IEC60335, GB4706.1-2005, GB4943 standard requirements
◆Measurement accuracy of electrical safety test is 3% and Measurement accuracy of power test is 0.5%
◆Can preset 15 groups, 7 items per group
◆Delete former forcibly self-check function, add compensation function to compensate GB, IR, ACW, LC test at any time, more convenient for users
◆Standard device for appliance test, suitable for testing more than 800 appliance production lines
◆Enhanced communication with remote control; test results can be saved instantly
◆Steady output supply voltage and current for all tests
◆Supply 50Hz and 60Hz frequency, a highly integrated electrical safety test instrument
◆Enhanced test analysis, memory, alarm setting, error offset, and auto protection functions
◆Equipped with customized high speed test programs
Technical Parameter | AN9651B(F) (6 in 1) | AN9641B(F) | AN9640B(F) | |||
Ground Bond Test | No-load Output Voltage | |||||
Current Presetting Range | AC 3A~30A 50Hz/60Hz 1A/step | AC 10A 25A 50Hz/60Hz | ||||
Current Accuracy | ±(1.5% of setting+0.40A) | |||||
Current Stability | ||||||
Resistance Range/Accuracy | 10~600mΩ ±(1.5% of reading+1.5% of range) | |||||
Resistance Up-Limit Range | 3A≤Current≤10A:10~600mΩ; 11A≤Current≤25A:10~300mΩ; | 10A:10~600mΩ; | ||||
Resistance Low-Limit Range | 0mΩ | |||||
Compensation Resistance Range | 0~200mΩ, Automatic Measurement, ON/OFF | |||||
Voltage Range/Accuracy | 30~7500mV ±(1.5% of reading+1.5% of range) | |||||
Voltage Up-Limit Range | 30~7500mV 1mV/step | |||||
Voltage Low-Limit Range | 0mV | |||||
Compensation Voltage Range | 0~1000mV, Automatic Measurement, ON/OFF | |||||
Test Time Range/Resolution | 1~999s 1s/step | |||||
Insulation Resistance Test | Rated Output Load/Current | 2VA(1000V/2mA) 2mA | ||||
Output Voltage Range/Resolution | DC 500V/1000V | |||||
Output Voltage Ripple | ||||||
Output Voltage Accuracy/Resolution | ±(3% of setting+3V) 5V | |||||
Resistance Range/Accuracy | ≤200MΩ: ±(3% of reading+3 counts); | |||||
Resistance Up-Limit Range | 0~2000MΩ | 2000MΩ | ||||
Resistance Low-Limit Range | 0.3~2000MΩ 0.1MΩ/step(≤100MΩ), 1MΩ/step(>100MΩ) | |||||
Test Time Range/Accuracy | 1~999s 1s/step | |||||
AC Hipot Test | Rated Output Load | 500VA(AC:5000V/100mA) | 100VA(AC:5000V/20mA) | |||
Output Voltage Range/Freq/Resolution | 500~5000V 50Hz/60Hz | |||||
Output Voltage Setting Range/Accuracy | Static:AC, 500~5000V;Dynamic:AC, 500~2500V; | Static:AC, 500~5000V; | ||||
Breakdown Current Meas. Accuracy | ±(3% of reading+0.03mA) | |||||
Breakdown Current Alarm Preset Range | Up-Limit:0.10-100.00mA;Low-Lmit:0-10.00mA; 0.01mA/step | Up-Limit:0.10-20.00mA;Low-Limit:0.0-10.00mA; 0.01mA/step | ||||
Compensation Current Range | 0.00-10.00mA, Automatic Measurement, ON/OFF | |||||
Test Time Range | 1~999s 1s/step | |||||
Rise Time Range | 0~100s 1s/step | |||||
Leakage Current Test | Output Voltage Range | 1.06 times of input voltage(output by isolated transformer) | external switching power or input by isolation mounting | |||
Leakage Mode Setting | Static, Dynamic | |||||
Output Voltage Meas. Range/Accuracy | AC 60~280V/45~65Hz; ±(0.5% of reading+2 counts) | |||||
Leakage Current Setting Range | Up-Limit:0~9.999mA 0.001mA;Low-Limit:0mA | |||||
Compensation Current Setting Range | 0~1000uA, Automatic Measurement, ON/OFF | |||||
Leakage Current Meas. Range/Accuracy | 0.050mA~9.999mA/ | |||||
Test Time Range | 1~999s 1s/step | |||||
Power Test | Output Voltage Range | 1.00 times of input voltage(output by isolated transformer) | External switching power or input by isolation mounting | |||
Power Setting Range/Resolution | Up-Limit:6~6000W Low-Limit:0~6000W 1W/step |
| Up-Limit:6~6000W Low-Limit:0~6000W 1W/step | |||
Output Voltage Meas. Range/Accuracy | AC 60~280V; 45~65Hz: ±(0.5% of reading+2 counts) |
| AC 60~280V; 45~65Hz: ±(0.5% of reading+2 counts) | |||
Current Meas. Range/Accuracy | AC 0.030~3.999A, 4.00~25.00A; |
| AC 0.030~3.999A, 4.00~25.00A; | |||
Real-Power Meas. Range/Accuracy | 10.0~199.9W: ±(0.5% of reading+5W) |
| 10.0~199.9W: ±(0.5% of reading+5W) | |||
Power Factor Meas. Range/Accuracy | 0.100~1.000; ±(2% of reading+2 counts) |
| 0.100~1.000; ±(2% of reading+2 counts) | |||
Test Time Range | 1~999s, 1s/step |
| 1~999s, 1s/step | |||
Low Voltage Start Test | Output Voltage Range | 1.00 times of input voltage(output by isolated transformer) |
|
| ||
Current Meas. Range/Resolution | Up-Limit:0.30~25.00A Low-Limit:0.00~25.00A 0.1A |
|
| |||
Output Voltage Meas. Range/Accuracy | AC 60~280V; 45~65Hz: ±(0.5% of reading+2 counts) |
|
| |||
Current Meas. Range/Accuracy | AC 0.030~3.999A, 4.00~25.00A |
|
| |||
Test Time Range | 1~999s 1s/step |
|
| |||
Display Mode | 240×128 LCD Large Screen | |||||
Isolated Transformer Capacity | Single Phase:2000W~6000W |
| ||||
Over-current Protection Limit | When current 30A, exceed 5s protection | |||||
Alarm Mode | Alarm with Sound and Light | |||||
Standard Interfaces | RS232 Interface, Alarm Light Interface, Remote Control Interface, AN97 Frequency Converter Power Supply Interface |
| ||||
Interfaces Optional | RS485 Interface, Printer Interface, Foot Pedal Interface, On-off Interface | RS485 Interface, Foot Pedal Interface, On-Off Interface | ||||
Buzzer Alarm Sound | Can ON and OFF | |||||
Test Group | 15 Group, 7 Test Steps/Group |
|